Imaging Document Consumer RAD-69 Retrieve Imaging Document Set:
Single Image Study
Purpose / Context
The Imaging Document Consumer performs:
RAD-69 Retrieve Imaging Document Set to retrieve the study for
the specified image.
The test points are:
The Imaging Document Consumer can perform a valid RAD-69
Retrieve Images independent of other transactions.
Test Steps
Setup
Images and KOS objects are pre-loaded into the testing system
actor simulators.
The Imaging Document Consumer will query the actor simulators
directly.
Instructions
The Imaging Document Consumer is instructed to query the
Registry/Repository simulator for the KOS object for the patient
with patient ID:
IDCAD013-a^^^&1.3.6.1.4.1.21367.2005.13.20.1000&ISO
The Imaging Document Consumer should both query for and
retrieve the KOS object.
The Imaging Document Consumer is instructed to use a RAD-69
Retrieve Imaging Document Set) transaction to retrieve the
imaging study with one image from the Imaging Document Source
simulator.
Validation
Test Manager: Detailed instructions on validating the query
and retrieve of the KOS are outlined in the test: ids_4831
'Retrieve KOS: Single Image Study'. Follow verification steps 1
through 6.
Test Manager: Verify that the retrieved KOS object is the
correct one. It should contain the following attributes: