RAD-69: SOAP Retrieve for Multi Image Study

Purpose / Context

The Imaging Document Source has submitted a KOS for a study with a single image. In this test, we retrieve the single image using a SOAP  retrieve.

Instructions

  1. Complete the instructions listed in test ids_2018-4801b. This will register the proper KOS object with the testing system. Assuming test ids_2018-4801b has completed successfully, you do not need to submit the data a second time.
  2. Make sure that the image data for patient IDS_2018-4801b^^^&1.3.6.1.4.1.21367.2005.13.20.1000&ISO is in your system and ready for a WADO retrieve.
  3. Activate the test button to initiate the SOAP retrieve and validation steps.

Notes

The software makes a single RAD-69 request for all of the images in the study. The single response contains all of the images in the study.

Validation

The test software provides the following validation tests:

  1. SOAP response is properly formatted.
  2. SOAP response contains proper content
  3. DICOM image is returned in SOAP response
  4. Original DICOM values in image remain unchanged
    1. Patient ID, Birth Date, Sex
    2. Study Instance UID
    3. Series Intance UID
    4. SOP Instance UID