Tests the ability of the Initiating Imaging Gateway actor (SUT) to respond correctly to a Retrieve Image Document Set (RAD-69) Request from an Image Document Consumer actor (Simulator), for a single DICOM image file, in the case where the Imaging Document Source has no image with matching DICOM UID values. The DICOM UIDs in the RAD-69 request do not refer to an image that is known to the Imaging Document Source (A1) behind the Responding Imaging Gateway (A), and an error code will be returned by the Responding Imaging Gateway to the Initiating Imaging Gateway.
RIG Home Community ID (A) | urn:oid:1.3.6.1.4.1.21367.13.70.101 |
IDS Repository Unique ID (A1) | 1.3.6.1.4.1.21367.13.71.101 |
Transfer Syntax UID | 1.2.840.10008.1.2.1 |
Note: Although the test environment provides for three Responding Imaging Gateways and multiple Image Document Sources, this test accesses a single Responding Imaging Gateway (A) and a single Imaging Document Source (A1) behind that gateway.
The test consists of three steps: